Ion Chromatography CHR15-042I
Chromatography is a technique that enables the separation, identification, and purification of the components of a mixture for qualitative and quantitative analysis. Our extensive range offers variety of products like Gas, Ion and Portable Ion chromatography products to meet all separation needs, including improved resolution, enhanced sensitivity, faster analysis and consistent performance.
Used in Food Testing, Chemical Industry, Beverage Testing, Drug testing, Forensic Science, Pharmaceutical, Molecular Biology, Medical, Research, Laboratory
Also known as Laboratory Chromatography
- Built-in circulating 3D constant temperature technology:
- Temperature stability time is less than 30 mins, ensuring the accuracy and
- reliability of test data.
- The world's leading full-range series of ion chromatographic columns:
Features
- Built-in circulating 3D constant temperature technology:
- Temperature stability time is less than 30 mins, ensuring the accuracy and
- reliability of test data.
- The world's leading full-range series of ion chromatographic columns:
- High efficiency, large capacity of the columns for detecting ions of varied
- compositions.
- Self-Regenerating Electrolytic Micro-membrane Suppressor:
- High pressure resistance, small dead volume, highly responsive to signals.
- Auto-range Conductivity Detector:
- It can directly detect the signal from ppb to ppm without adjusting the range. Only one conductivity detector can detect anions and cations.
- Observatory intelligent workstation:
- Integrated control, intelligent start-up, shutdown and maintenance functions. Compatible with a variety of instruments.
Specifications
Ion Chromatographic Pump: | |
Type | Self-Regenerating electrolytic micro-membrane suppressor |
Maximum Pressure | 6.0 Mpa |
Pressure Display Accuracy | ≤ 0.1 MPa |
Flow Range | 0.001 ~ 9.999 mL/min |
Pressure Fluctuation | ≤ 0.5% |
Flow Stability | (0.2-0.5) mL/min ≤ 2%; (0.5-1.0) mL/min ≤ 1%; > 1.0 mL/min ≤ 1% |
Manual Sample Injector: | |
Contact Material of the Rotor | PEEK |
Contact Material of Medium | PEEK/Ceramics |
Column Heater: | |
Operating Temperature Range | Room temperature +5°C~60°C(41~140°F) |
Controlling Temperature Accuracy | ±0.01°C |
Allowable Deviation of Column Heater’s Temperature | ± 1°C |
Temperature Stability | ≤ 0.05°C/h |
Conduction Detection System: | |
Cell Volume | ≤0.8μL |
Detection Range | 0~35000 μS/cm |
Detection Resolution | ≤0.0020nS/cm |
Output Voltage | -6000~+6000 mv (adjustable) |
Electronic Noise | 0.02 nS |
Baseline Noise | ≤ 0.001 μS/cm |
Baseline Drift | ≤ 0.02μS |
Temperature Compensation | 1.7 %/°C |
Linear Range | ≥ 10 3 |
Instrument Linearity | ≥0.999 |
Quantitative Repeatability | ≤1.0% |
Qualitative Repeatability | ≤0.5% |
Minimum Detectable Concentration | Cl- ≤ 0.0005 ug/mL; Li+ ≤ 0.001 ug/mL; BrO3≤ 0.001 ug/mL |
Flow System: | |
Six-way Valve | PEEK material, pressure 5000 psi; Independent automatic collecting and flow function. |
Suppressor: | |
Dead Volume | <50 μL |
Other Specifications: | |
Dimension (LxWxH) | 350x470x510 mm |
Net Weight | 26 kg |
Gross Weight | 32 kg |
Power | 150 W |