Home /
Water Purification System
/ Ultrapure Water Purification System
/ Standard Ultrapure Water System WPS61-030R
Standard Ultrapure Water System WPS61-030R
Ultrapure water system is sub-economic choice for high grade experiments. This level of purification is required for advanced analytical techniques, such as HPLC, and is commonly used for semi-conductor manufacturing.
Used in Laboratory, Manufacturing, Reefkeeping, Aquarium, Laboratory, Research
Also known as Laboratory Ultrapure water system
- With tap water inlet, to produce RO water and ultrapure water,quality can reach to above10MΩ.cm.
- Built-in 20 liters airtight plastic pressure water tank
- Built-in 13 liters high-capacity polishing resin cartridge
- Unique design and easy-to-replace cartridges pack unit.
Features
- With tap water inlet, to produce RO water and ultrapure water,quality can reach to above10MΩ.cm.
- Built-in 20 liters airtight plastic pressure water tank
- Built-in 13 liters high-capacity polishing resin cartridge
- Unique design and easy-to-replace cartridges pack unit.
- Data storage and RS 232/USB communication port.
- 3 way on-line water quality sensor,multiple alarm.
- Life-span of cartridges´ display and alarm.
- System circulation function,system sterilization procedure.(optional)
- Molding process,high-strength, streamline plastic shell.
- The graphic display clearly indicates all system´s parameters. From water quality to knowing when it is time to change the purification pack,you´ll see at a glance what is need
- For ease-of-use,the main purification technologies are contained in an innovative all-in-one pack that mean you can change it in just a couple of minutes.
- The system requires no special installation,connect the system to your tap water supply it´s ready to use.
Specifications
Feed Water Requirements*: | |
Water Inlet | Tap water: TDS<200 ppm (Extra pretreatment filter is recommended, if TDS>200 ppm) |
Temperature | 5-45°C |
Pressure | 1.0-4.0 Kgf/cm² |
Flow Procedure** | PF+AC+RO+DI+TF |
Ion rejection rate | 96%-99% (New RO membrane) |
Organic rejection rate | >99%,when MW>200 Dalton |
Particles and bacteria rejection rate | >99% |
Bacteria | <0.1 cfu/ml |
Output(25°C)**** | 30 L/hrs |
Pure water outlet | RO water and Ultrapure water |
Water Quality Monitor | Portable TDS/conductivity test pen + on-line resistivity monitor |
DimensionLxWxH | 410x220x420 mm |
Weight | 20 kg |
Standard configuration | Main body (Including 1 set of cartridges)+15 liters tank+ TDS pen +accessory bag |
Power Consumption (W) | 72 W |
Power Supply | AC110-220 V, 50/60 Hz |
Note | *The feed water quality will influence the pure water's quality and cartridges life-span. **PF:polypropylene spun fiber, AC:active carbon, RO:reverse osmosis, DI:ion exchange, UV:ultraviolet, UF:ultrafiltration, TF:terminal microfiltration. ***Value of number will be influenced by temperature and feed water quality. ****All the specifications are tested under the situation:feed water's TDS=200ppm, 25°C, 50psi and 15% recovery rate. |
Deionized water quality: | |
Particle(>0.2µm) | <1/ml |
Ultrapure Water Quality: | |
Resistivity(25°C) | 18.2 MΩ.cm |
Heavy Metal Ion | - |
TOC*** | <10 ppb |
Heavy metal ion | <0.1 ppb |
Flow rate | 2.0 L/min (with pressure tank) |